PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-24569
- Event Type
- Malfunction
- Date Received
- September 18, 2014
- Date of Event
- August 19, 2014
- Report Date
- August 19, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED DURING TESTING. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, CRACKED BATTERY TUBE THREADS, A CRACKED RESERVOIR TUBE LIP AND A MISSING END CAP STICKER.
THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC (B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 215 MG/DL. NO FURTHER INFORMATION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 579418 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | OYC | MEDTRONIC MINIMED | MMT-754WWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |