ANIMAS VIBE
Report
- Report Number
- 2531779-2014-27663
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Report Date
- September 12, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(4).
FOLLOW-UP #1: DATE OF SUBMISSION (B)(4) 2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: INVESTIGATION REVEALED EVIDENCE OF MOISTURE IN THE BATTERY COMPARTMENT, IN THE DISPLAY, AND ON THE PCB. THE PUMP PASSED LEAK TESTING AND THERE WAS NO DAMAGE FOUND TO THE PUMP CASING. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THERE WAS MOISTURE IN THE PUMP. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 600609 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |