FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 4122381 · Received September 26, 2014

Report

Report Number
1416980-2014-33412
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
June 10, 2014
Report Date
September 3, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION WAS COMPLETED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASE INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A HIGH DRAIN ERROR 106 (NIGHT DRAIN #6) ALARM WAS IDENTIFIED IN THE LOG. THIS INDICATES THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME IN STANDARD MODE. THE ALARM OCCURRED ON (B)(6) 2014 AT 13:02:35. NO FURTHER INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600608 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment