JAGWIRE?
Report
- Report Number
- 3005099803-2014-03225
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- September 4, 2014
- Report Date
- September 5, 2014
- Manufacturer
- BOSTON SCIENTIFIC - COSTA RICA (COYOL)
- Product Code
- EZB
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CN
- Reporter Occupation
- PHYSICIAN
Narratives
A VISUAL EXAMINATION OF THE GUIDEWIRE PRESENTED NO DAMAGE. NO EVIDENCE OF COREWIRE FRACTURE AND THE PTFE JACKET DID NOT PRESENT ANY ANOMALY. THE COMPLAINT IS NOT CONSISTENT WITH THE RETURN THAT THE DISTAL TIP WAS DETACHED EXPOSING THE TIP OF THE METAL COREWIRE. THE DEVICE SHOWED NEITHER EVIDENCE OF THE ALLEGED ISSUE, NOR ANY DEFECT WHICH COULD HAVE CONTRIBUTED TO THE REPORTED EVENT, THEREFORE, THE MOST PROBABLE ROOT CAUSE IS NOT CONFIRMED. A DHR (DEVICE HISTORY RECORD) REVIEW WAS PERFORMED AND NO DEVIATION WAS FOUND.
(B)(4). THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE GUIDEWIRE WAS USED DURING A CHOLEDOCHOLITHIASIS TREATMENT PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE HYDROPHILIC TIP OF THE JAGWIRE GUIDEWIRE WAS DETACHED INSIDE THE HEPATIC DUCT EXPOSING THE TIP OF THE METAL COREWIRE. THE PHYSICIAN ATTEMPTED TO RETRIEVE THE DETACHED TIP BUT FAILED. THE PHYSICIAN IS MONITORING THE PATIENT TO SEE IF THE DETACHED TIP WILL PASS NATURALLY THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO SERIOUS INJURIES REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE GUIDEWIRE WAS USED DURING A CHOLEDOCHOLITHIASIS TREATMENT PROCEDURE PERFORMED ON (B)(6), 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE HYDROPHILIC TIP OF THE JAGWIRE GUIDEWIRE WAS DETACHED INSIDE THE HEPATIC DUCT EXPOSING THE TIP OF THE METAL COREWIRE. THE PHYSICIAN ATTEMPTED TO RETRIEVE THE DETACHED TIP BUT FAILED. THE PHYSICIAN IS MONITORING THE PATIENT TO SEE IF THE DETACHED TIP WILL PASS NATURALLY THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO SERIOUS INJURIES REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 600892 | JAGWIRE? | STYLET FOR CATHETER, GASTRO-UROLOGY | EZB | BOSTON SCIENTIFIC - COSTA RICA (COYOL) | M0055658011 | 16293728 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR |