FDA Adverse Event Malfunction Summary report: N

14CM ANGLE ATTACHMENT

MDR report key: 4122316 · Received September 26, 2014

Report

Report Number
1045834-2014-13090
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
July 19, 2013
Report Date
July 19, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO WORN BEARINGS FROM NORMAL USE AND SERVICING OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED FROM THE NETHERLANDS THAT UPON RECEIPT OF A NEWLY PURCHASED ATTACHMENT DEVICE AT THE HOSPITAL, IT WAS OBSERVED THAT IT LOOKED USED AND WORN; AND THAT THE DRIVER SHAFT SHOWED SIGNS OF WEAR. DURING IN-HOUSE ENGINEERING EVALUATION AT PRE-TESTING, IT WAS DISCOVERED THAT THE DEVICE WAS FOUND TO HAVE A HIGH TEMPERATURE. THE EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO INJURIES OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
601238 14CM ANGLE ATTACHMENT MOTOR, DRILL, ELECTRIC - ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1