FDA Adverse Event Summary report: N

20" CENTURY VACUUM STEAM STERILIZER

MDR report key: 4122270 · Received September 26, 2014

Report

Report Number
3005899764-2014-00093
Date Received
September 26, 2014
Date of Event
September 3, 2014
Report Date
September 26, 2014
Manufacturer
STERIS MEXICO, S. DE R.L. DE C.V.
Product Code
FLE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS FIELD SERVICE TECHNICIAN ARRIVED ONSITE AND INSPECTED THE UNIT. THE TECHNICIAN IDENTIFIED THAT THE CAUSE OF THE LEAK WAS DUE TO A WORN DRAIN FUNNEL. THE TECHNICIAN REPLACED THE DRAIN FUNNEL, TESTED THE STERILIZER, AND CONFIRMED THE UNIT WAS OPERATING ACCORDING TO SPECIFICATION. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Description of Event or Problem · 1

THE USER FACILITY REPORTED A LEAK FROM THEIR CENTURY STERILIZER. NO INJURIES OR PROCEDURAL DELAYS/CANCELLATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600866 20" CENTURY VACUUM STEAM STERILIZER STERILIZER FLE STERIS MEXICO, S. DE R.L. DE C.V.

Patients

Seq Age Sex Outcome Treatment
1