FDA Adverse Event Malfunction Summary report: N

6 SHOOTER SAEED MULTI-BAND LIGATOR

MDR report key: 4122254 · Received September 17, 2014

Report

Report Number
1037905-2014-00358
Event Type
Malfunction
Date Received
September 17, 2014
Report Date
August 18, 2014
Manufacturer
WILSON-COOK MEDICAL INC.
Product Code
MND
PMA / PMN Number
K944220
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. THE LOT NUMBER ASSOCIATED WITH THIS COMPLAINT WAS RESEARCHED TO DETERMINE THE MANUFACTURING DATE OF THE ACTUAL BANDS. WE CAN CONCLUDE FROM THESE DATES THAT THE BANDS WERE WITHIN THE ACCEPTABLE AGE DATE RANGE TO ENSURE THE BANDS MAINTAIN THEIR ELASTICITY PROPERTIES. THE TRIGGER CORD SUB ASSEMBLY DEVICE HISTORY RECORD WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. RAPID ROTATION OF THE LIGATOR HANDLE CAN CONTRIBUTE TO PREMATURE BAND DEPLOYMENT. THE INSTRUCTIONS FOR USE DIRECT THE USER TO MAINTAIN SUCTION AND DEPLOY THE BAND BY ROTATING THE HANDLE CLOCKWISE UNTIL BAND RELEASE IS FELT. DURING THE PROCEDURE, ENDOSCOPIC SUCTION IS APPLIED TO THE BANDING SITE TO PROPERLY PLACE A LIGATOR BAND. PREMATURE BAND DEPLOYMENT CAN ALSO OCCUR IF THE HANDLE IS ROTATED BEFORE MAINTAINING SUCTION ON THE BANDING SITE. THE INSTRUCTIONS FOR USE ADVISE THE USER TO MAINTAIN SUCTION WHILE DEPLOYING THE BAND. IF THE LIGATOR HANDLE IS IN THE FIRING POSITION WHILE THE ENDOSCOPE IS STRAIGHTENED, THIS CAN RESULT IN PREMATURE BAND DEPLOYMENT DUE TO TENSION ON THE TRIGGER CORD. THE INSTRUCTIONS FOR USE DIRECT THE USER TO PLACE THE HANDLE IN THE TWO-WAY POSITION BEFORE STRAIGHTENING THE ENDOSCOPE. ANOTHER POSSIBLE CONTRIBUTING FACTOR TO PREMATURE BAND DEPLOYMENT INCLUDES ALLOWING THE TRIGGER CORD TO BECOME LODGED BETWEEN THE BARREL AND THE DISTAL END OF THE ENDOSCOPE. THIS CAN RESTRICT TRIGGER CORD MOVEMENT AND RESULT IN PREMATURE BAND DEPLOYMENT IF ENOUGH TENSION IS APPLIED TO PULL THE CORD FREE. PRIOR TO DISTRIBUTION, ALL 6 SHOOTER SAEED MULTI-BAND LIGATORS ARE SUBJECTED TO A VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 1

DURING THE ESOPHAGOGASTRODUODENOSCOPY (EGD), A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR WAS USED TO BAND ESOPHAGEAL VARICES. MULTIPLE BANDS WERE DEPLOYING AT ONE TIME. ANOTHER OF THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. OTHER THAN THE DEPLOYED BANDS, A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE AN ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
575597 6 SHOOTER SAEED MULTI-BAND LIGATOR MND, LIGATOR, ESOPHAGEAL MND WILSON-COOK MEDICAL INC. W3434221

Patients

Seq Age Sex Outcome Treatment
1 OLYMPUS ENDOSCOPE (UNK MODEL NUMBER)