FDA Adverse Event Malfunction Summary report: N

DPM 6 MONITOR

MDR report key: 4122202 · Received September 18, 2014

Report

Report Number
2221819-2014-00548
Event Type
Malfunction
Date Received
September 18, 2014
Date of Event
July 23, 2014
Report Date
July 30, 2014
Manufacturer
MINDRAY BIOMEDICAL ELECTRONICS CO. LTD
Product Code
MHX
PMA / PMN Number
092449
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

COMPANY REPRESENTATIVE EVALUATED THE UNIT. CORRECTIONS INCLUDED REPLACEMENT OF THE UNIT'S MPM MODULE FRONT PANEL. UNIT WAS CALIBRATED AND TESTED TO FACTORY'S SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED AN ISSUE WITH THE DPM 6 MONITOR, WHICH MAY HAVE AFFECTED PATIENT MONITORING. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
579061 DPM 6 MONITOR PATIENT MONITOR MHX MINDRAY BIOMEDICAL ELECTRONICS CO. LTD

Patients

Seq Age Sex Outcome Treatment
1