FDA Adverse Event Malfunction Summary report: N

SYMBIQ 3.13 SINGLE C

MDR report key: 4122008 · Received September 19, 2014

Report

Report Number
9615050-2014-05322
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
January 1, 2014
Report Date
September 3, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0069-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING THE DEVICE TOUCHSCREEN WAS FOUND TO BE OUT OF CALIBRATION. THE TOUCH POINTS ON THE TOUCHSCREEN WERE OFF CENTER. THE PROBABLE CAUSE WAS A BROKEN TOUCHSCREEN. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN WOULD NOT CALIBRATE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FROM INTENSIVE CARE. IT WAS INDICATED THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE TOUCHSCREEN WAS CUT OF CALIBRATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
583631 SYMBIQ 3.13 SINGLE C 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA