FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 4121724
·
Received September 26, 2014
Report
- Report Number
- 1416980-2014-33366
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- September 2, 2014
- Report Date
- September 2, 2014
- Manufacturer
- SHARP CORPORATION OSAKA JAPAN
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE PERFORMED. THE DAMAGED BATTERY WAS IDENTIFIED DURING VISUAL TESTING. THE CAUSE OF THE CONDITION WAS NOT DETERMINED. TO CORRECT THE CONDITION, THE MAIN BATTERY WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
DURING PRODUCT SERVICE BY A SERVICE TECHNICIAN, A FLOGARD INFUSION PUMP WAS FOUND TO HAVE A DAMAGED BATTERY. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 601982 | FLOGARD | PUMP, INFUSION | FRN | SHARP CORPORATION OSAKA JAPAN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |