FDA Adverse Event
Malfunction
Summary report: N
POSEY KEEPSAFE DELUXE
MDR report key: 4121591
·
Received July 29, 2014
Report
- Report Number
- 2020362-2014-00251
- Event Type
- Malfunction
- Date Received
- July 29, 2014
- Report Date
- June 27, 2014
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
RESULTS: EVALUATION OF THE RETURNED PRODUCT CONFIRMED THE REPORTED ISSUE. THE UNIT'S VOICE MESSAGE DOES NOT PLAY WHEN WEIGHT IS REMOVED FROM THE SENSOR PAD, ONLY A CLICKING SOUND IS HEARD. THE UNIT'S TONE PLAYS PROPERLY. THERE IS BATTERY LEAKAGE AND CORROSION ON THE BATTERY FLAT CONTACTS. (B)(4).
Description of Event or Problem · 1
CUSTOMER REPORTED THAT WHEN THE ALARM IS SET TO VOICE AND TONE MODE, THE PRE-RECORDED MESSAGE IS NOT HEARD. THE REPORTED ISSUE WAS DISCOVERED DURING SET UP, BUT THE DATE OF THE EVENT IS UNKNOWN. NO PATIENT INCIDENT OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 442715 | POSEY KEEPSAFE DELUXE | KMI | J. T. POSEY CO. | 8374 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CHAIR SENSOR, SQUARE, 6-MONTH, MODEL 8308,| LOT # UNKNOWN. |