FDA Adverse Event
Malfunction
Summary report: N
TENDRIL SDX
MDR report key: 4121511
·
Received January 13, 2014
Report
- Report Number
- 2017865-2014-07752
- Event Type
- Malfunction
- Date Received
- January 13, 2014
- Date of Event
- July 5, 2012
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- DTB
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ATRIAL LEAD EXHIBITED LOSS OF CAPTURE, LOSS OF SENSING, AND LEAD FRACTURE. IT WAS RESOLVED BY DEACTIVATING LEAD. THE LEAD REMAINED IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 32256 | TENDRIL SDX | PERMANENT PACEMAKER ELECTRODE | DTB | ST. JUDE MEDICAL, INC., CRMD | 1688TC/52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |