FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 4121478 · Received September 26, 2014

Report

Report Number
1031452-2014-11140
Event Type
Malfunction
Date Received
September 26, 2014
Report Date
September 8, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

REPAIR CENTER: ALLEGED LOW O2/YELLOW LIGHT AND KEY IS THE VALVE IS THE MANIFOLD VALVE STUCK /LEAKING. ADDITIONAL MALFUNCTIONS ARE THE INLET FILTER IS DIRTY, THE PE VALVE IS LEAKING, AND THE ZIP TIES AND HOSE CLAMPS WERE REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
601428 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW INVACARE FLORIDA OPERATIONS IRC5PO2V

Patients

Seq Age Sex Outcome Treatment
1 Other