FDA Adverse Event
Malfunction
Summary report: N
PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 4121452
·
Received September 26, 2014
Report
- Report Number
- 1031452-2014-11138
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Report Date
- September 8, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
PER THE REPAIR CENTER: ALLEGED LOW O2/YELLOW LIGHT AND KEY FAILURE WAS THE MANIFOLD VALVE LEAKING, AND STICKING. ADDITIONAL MALFUNCTIONS ARE THE INLET FILTER WAS DIRTY AND THE ZIP TIES AND HOSE CLAMPS WERE REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 601989 | PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) | 868.5440 | CAW | INVACARE FLORIDA OPERATIONS | IRC5PO2V |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |