ACCENT DR RF
MDR report key: 4121397
·
Received January 13, 2014
Report
- Report Number
- 2017865-2014-06984
- Event Type
- Malfunction
- Date Received
- January 13, 2014
- Date of Event
- April 4, 2011
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- DXY
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD. EVALUATION: FINAL ANALYSIS COULD NOT CONFIRM CAPTURE ANOMALIES. FLUID WAS FOUND IN THE ATRIAL AND VENTRICULAR CONNECTORS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WAS EXHIBITING CAPTURE ANOMALIES WITH VARYING PACING THRESHOLDS. FLUID WAS NOTED IN THE HEADER. THE DEVICE WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 32445 | ACCENT DR RF | IMPLANTABLE PACEMAKER PULSE GENERATOR | DXY | ST. JUDE MEDICAL, INC., CRMD | PM2210 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR |