FDA Adverse Event Malfunction Summary report: N

ACCENT DR RF

MDR report key: 4121397 · Received January 13, 2014

Report

Report Number
2017865-2014-06984
Event Type
Malfunction
Date Received
January 13, 2014
Date of Event
April 4, 2011
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
DXY
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD. EVALUATION: FINAL ANALYSIS COULD NOT CONFIRM CAPTURE ANOMALIES. FLUID WAS FOUND IN THE ATRIAL AND VENTRICULAR CONNECTORS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS EXHIBITING CAPTURE ANOMALIES WITH VARYING PACING THRESHOLDS. FLUID WAS NOTED IN THE HEADER. THE DEVICE WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
32445 ACCENT DR RF IMPLANTABLE PACEMAKER PULSE GENERATOR DXY ST. JUDE MEDICAL, INC., CRMD PM2210

Patients

Seq Age Sex Outcome Treatment
1 78 YR