HOMECHOICE PRO
Report
- Report Number
- 1416980-2014-33330
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- August 18, 2014
- Report Date
- September 1, 2014
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS COMPLETED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED IN THE EVENT HISTORY LOG REVIEW. THE CAUSE OF THE IIPV EVENT WAS ONE OR MORE CYCLES ADVANCED TO THE NEXT FILL WHEN A SLOW OR NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 19:29:44. DURING NIGHT DRAIN CYCLE TWO, THE PATIENT'S ULTRAFILTRATION READING WAS 1204ML, INDICATING THE HOME PATIENT (HP) DRAINED 1204ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.