FLOGARD
MDR report key: 4121195
·
Received September 26, 2014
Report
- Report Number
- 1416980-2014-33302
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- September 1, 2014
- Report Date
- September 1, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. THE DEVICE WAS VISUALLY INSPECTED, FUNCTIONALLY TESTED AND AN ALARM LOG REVIEW WAS PERFORMED. DURING SERVICE, THE BATTERY WAS FOUND TO BE SWOLLEN; THE CAUSE WAS UNDETERMINED. TO CORRECT THE CONDITION, THE BATTERY WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
DURING PRODUCT SERVICE BY A SERVICE TECHNICIAN, A FLOGARD INFUSION PUMP HAD A SWOLLEN BATTERY. NO ADDITIONAL INFORMATION IS AVAILABLE.