FDA Adverse Event Malfunction Summary report: N

SENSOR ENLITE

MDR report key: 4121016 · Received September 26, 2014

Report

Report Number
2032227-2014-29405
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
August 28, 2014
Report Date
August 28, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD NORMAL BLOOD GLUCOSE LEVELS IN THE RANGE OF 85 MG/DL TO 115 MG/DL. THE CUSTOMER REPORTED HAVING ISSUES WITH THE SENSOR GLUCOSE LEVELS READING BEING DIFFERENT FROM THE BLOOD GLUCOSE LEVELS READING. THE CUSTOMER ALSO REPORTED THAT THE THRESHOLD SUSPEND FEATURE OF THE INSULIN PUMP ACTIVATED EVEN THOUGH IT WAS NOT SUPPOSED TO. THE HEALTH CARE PROVIDER OF THE CUSTOMER ADJUSTED THE BASAL SETTINGS OF THE CUSTOMER TO AVOID THE THRESHOLD SUSPEND ACTIVATING INAPPROPRIATELY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
601614 SENSOR ENLITE OZO OZO MEDTRONIC MINIMED MMT-7008A F044U

Patients

Seq Age Sex Outcome Treatment
1 67 YR