PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-28889
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- August 28, 2014
- Report Date
- August 28, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP POWERED UP PROPERLY AFTER BATTERY INSTALLATION. THE INSULIN PUMP HAD OPERATING CURRENTS WITHIN SPECIFICATION AND NO UNEXPECTED BATTERY OUT LIMIT ALARMS WERE NOTED. THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED. NO DISPLAY ANOMALY WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP AND CRACKED BATTERY TUBE THREADS.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BATTERY OUT LIMIT, AND WHEN THE CUSTOMER ATTEMPTED TO CLEAR THE ALARM, THE DEVICE HAD AN UNRESPONSIVE KEYPAD. THE BLOOD GLUCOSE READING WAS 133 MG/DL. NO SIGNIFICANT EVENTS LEADING TO THE KEYPAD ISSUES WERE OBSERVED. ADVISED REPLACEMENT OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 601525 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-522LNAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |