FDA Adverse Event Malfunction Summary report: N

THE EMAX 2 PLUS SYSTEM

MDR report key: 4120879 · Received September 26, 2014

Report

Report Number
1045834-2014-13055
Event Type
Malfunction
Date Received
September 26, 2014
Report Date
January 15, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK080802
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY TECHNICIAN, INVESTIGATION REVEALED THE DEVICE HOUSING WAS DAMAGED AND THE HOSE WAS MISSING THE PLUG. THERE WAS NO EVIDENCE OF DEVICE OVERHEATING DURING THE INVESTIGATION. THE DEVICE WAS REPAIRED AND RETURNED TO CUSTOMER ON (B)(4) 2013. MFR DATE: 8/29/2011.

Description of Event or Problem · 1

CUSTOMER REPORTED THE DEVICE OVERHEATS AFTER 3-5 MINUTES AND MAKES EXCESSIVE NOISE. THIS IS REPORT 1 OF 1 FOR #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
601994 THE EMAX 2 PLUS SYSTEM MOTOR, DRILL, ELECTRIC HBC DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1