FDA Adverse Event Malfunction Summary report: N

TIP FOR DHS®/DCS® IMPACTOR

MDR report key: 4120631 · Received September 26, 2014

Report

Report Number
2530088-2014-10287
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
June 24, 2014
Report Date
June 24, 2014
Manufacturer
SYNTHES BRANDYWINE
Product Code
HWA
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(6). DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. A DEVICE HISTORY REVIEW WAS CONDUCTED. THE REPORT INDICATES THAT THERE WERE NO ISSUES WERE FOUND DURING MANUFACTURE THAT WOULD CONTRIBUTE TO THE COMPLAINT CONDITION. A MANUFACTURING INVESTIGATION WAS CONDUCTED. THE REPORT INDICATES THAT JUDSON A. SMITH, CO. MANUFACTURED THE TIP FOR DHS/DCS IMPACTOR, P/N 338.26, AND LOT NUMBER 4550661 ON PO 356297 DATED MARCH 26, 2003. DUE TO AN UNKNOWN CAUSE, THE TIP BROKE AND BROKEN PIECE(S) ARE NOT WITH THE COMPLAINT. INITIALLY, THE PART CONFORMED TO THE CERTIFICATE OF COMPLIANCE (DATED MARCH 24, 2003 FOR 48 PIECES) AND WAS INSPECTED AND CONFORMED TO THE SYNTHES INCOMING FINAL INSPECTION SHEET #MCN0000, REVISION ¿L¿. BASED ON THE EVALUATION AND THE UNKNOWN CAUSE, THIS COMPLAINT IS NOT MANUFACTURING-RELATED AND IS CONSIDERED UNCONFIRMED. THE TIP FOR DHS/DCS IMPACTOR ASSEMBLY WAS MADE TO THE SYNTHES DRAWING P/N 338.26, REVISION ¿C¿, RELEASED ON 10/15/1997 DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SURGERY THE SURGEON WAS USING THE IMPACTOR TO IMPACT A PLATE AND THAT THE TIP OF THE IMPACTOR BROKE. THERE WAS A BACKUP TO COMPLETE THE CASE AND THERE WAS NO SURGICAL DELAY. SURGERY WAS SUCCESSFULLY COMPLETED WITH GOOD OUTCOME. NO PATIENT HARM. THIS IS REPORT NUMBER 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600849 TIP FOR DHS®/DCS® IMPACTOR IMPACTOR HWA SYNTHES BRANDYWINE 4550661

Patients

Seq Age Sex Outcome Treatment
1 45 YR