FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 4120562 · Received September 26, 2014

Report

Report Number
1031452-2014-11063
Event Type
Malfunction
Date Received
September 26, 2014
Report Date
September 5, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

PER INDEPENDENT REPAIR CENTER STATEMENT, UNIT WAS LOW O2 OR YELLOW LIGHT. THE KEY FAILURE WAS THAT THE PILOT VALVE IN THE MANIFOLD WAS LEAKING. ADDITIONAL MALFUNCTIONS ARE THE O-RING FOR THE MANIFOLD AND THE TIE STRAPS FOR THE SIEVE BEDS WERE DEFECTIVE AND THE SOUND BOX FILTER AND CABINET FILTER WERE MISSING.