FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 4120557 · Received September 26, 2014

Report

Report Number
1031452-2014-11058
Event Type
Malfunction
Date Received
September 26, 2014
Report Date
September 5, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

INDEPENDENT REPAIR CENTER: ALARMING OR RED LIGHT. KEY FAILURE IS PE VALVE LEAKING. ADDITIONAL MALFUNCTIONS ARE THE POWER SWITCH HAS A SHORT CIRCUIT, THE FILTER FOR THE SOUND BOX IS MISSING, AND THE O-RING FOR THE MANIFOLD AND TIE STRAPS FOR THE SIEVE BEDS ARE DEFECTIVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600741 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW INVACARE FLORIDA OPERATIONS IRC5PO2V

Patients

Seq Age Sex Outcome Treatment
1 Other