HOMECHOICE
Report
- Report Number
- 1416980-2014-33288
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- September 2, 2014
- Report Date
- September 2, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HAS BEEN RECEIVED AND THE EVALUATION IS IN PROGRESS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED TO INVESTIGATE THE REPORTED PROBLEM. A REVIEW OF THE EVENT HISTORY LOG DID NOT IDENTIFY ISSUES RELATED TO THE REPORTED PROBLEM. A REVIEW OF THE SERVICE HISTORY REVEALED NO PREVIOUS SERVICE EVENTS THAT WOULD CAUSE OR CONTRIBUTE TO THE REPORTED PROBLEM. AN INTERNAL AND EXTERNAL VISUAL INSPECTION REVEALED NO PROBLEMS. THE HOMECHOICE DEVICE RECEIVED A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDED FUNCTIONAL AND ELECTRICAL TESTING OF THE DEVICE. THE DEVICE PASSED THE ELECTRICAL TESTING BUT FAILED THE RITE FUNCTIONAL TESTING FOR RITE - THERAPY MONITORED VOLUME FAILED PERFORMANCE SPEC. THE DIRECT CAUSE OF THIS PROBLEM WAS DETERIORATED PISTON FOAM. THE DETERIORATED PISTON FOAM WAS SCRAPPED TO SOLVE THE PROBLEM. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED FLUID VOLUME ACCURACY TESTING. DEVICE FAILED DURING EVALUATION, NO PATIENT INVOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 600995 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |