FDA Adverse Event Malfunction Summary report: N

IMPACTOR ASSEMBLY OMEGA

MDR report key: 4120502 · Received September 26, 2014

Report

Report Number
0008031020-2014-00446
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
September 3, 2014
Report Date
September 3, 2014
Manufacturer
STRYKER GMBH
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED EVENT OF THE IMPACTOR BREAKAGE COULD BE CONFIRMED, SINCE THE RETURNED DEVICE MATCHED THE REPORTED FAILURE. BASED ON THE INVESTIGATION, THE ROOT CAUSE WAS ATTRIBUTED TO A USER RELATED ISSUE. THE BREAKAGE WAS CAUSED BY THE APPLICATION OF EXCESSIVE FORCE ON THE DEVICE. THE VISUAL INSPECTION REVEALED THAT THE BREAKAGE IS CONSISTENT WITH THE APPLICATION OF A HIGH FORCE ON THE DEVICE. NOTE, AS STATED IN THE OPERATIVE TECHNIQUE: ¿IMPACTION OF THE FRACTURE MAY BE ACCOMPLISHED BY USING THE PLATE IMPACTOR TOGETHER WITH A HAMMER OR MALLET. NOTE: USE GENTLE HAMMERING ONLY - OTHERWISE THE IMPACTOR MAY BE DESTROYED.¿ A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. NO INDICATIONS OF MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE FOUND DURING THE INVESTIGATION.

Description of Event or Problem · 1

DURING THE INSPECTION OF THE RETURNED LONER INSTRUMENTS FROM THE HOSPITAL, IT WAS FOUND THAT THE IMPACTOR BROKE. ACCORDING TO SALES REP, DURING SURGERY WHEN THE SURGEON USED THE HAMMER, THE IMPACTOR BROKE. ALL FRAGMENTS WERE DISCARDED IN THE HOSPITAL.

Description of Event or Problem · 1

DURING THE INSPECTION OF THE RETURNED LONER INSTRUMENTS FROM THE HOSPITAL, IT WAS FOUND THAT THE IMPACTOR BROKE. ACCORDING TO SALES REP, DURING SURGERY WHEN THE SURGEON USED THE HAMMER, THE IMPACTOR BROKE. ALL FRAGMENTS WERE DISCARDED IN THE HOSPITAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600631 IMPACTOR ASSEMBLY OMEGA ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LXH STRYKER GMBH D00501

Patients

Seq Age Sex Outcome Treatment
1 Other