ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-27547
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Report Date
- September 15, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX HISTORY SHOWED NO EVIDENCE OF ¿NO POWER¿ EVENTS OCCURRING. THE BATTERY COMPARTMENT WAS INTACT; NO DAMAGE WAS OBSERVED. NO BATTERY CAP WAS RETURNED WITH THE PUMP. A TEST BATTERY CAP WAS ABLE TO BE FULLY SECURED TO THE PUMP AND WAS USED TO COMPLETE ALL TESTING. DURING TESTING, THE PUMP WAS EXERCISED FOR 24 HOURS ON A 1 UNIT PER HOUR BASAL RATE WITH NO POWER REBOOTS, POWER LOSSES, OR CALL SERVICE ALARMS OCCURRING. THE PUMP CASE WAS REMOVED AND NO EVIDENCE OF MOISTURE OR LOOSE COMPONENTS WAS FOUND INSIDE THE PUMP. THE COMPLAINT THAT THE PUMP HAD NO POWER WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION. NO DEFECTS WERE FOUND.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (NO POWER) ISSUE. THE REPORTER ALLEGED THAT THE PUMP HAD NO POWER. IT WAS NOTED THAT THE POWER ISSUE OCCURRED DURING OR IMMEDIATELY AFTER CHANGING THE BATTERY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 601648 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |