ANIMAS INSULIN INFUSION PUMP

MDR report key: 4120302 · Received September 26, 2014

Report

Report Number
2531779-2014-27535
Event Type
Malfunction
Date Received
September 26, 2014
Report Date
September 15, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING, THE DISPLAY WAS FOUND TO BE DIM, FADED, AND DISCOLORED. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED BELOW THE BUMPER PAD. ALSO UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THAT THE UP ARROW, DOWN ARROW, AND OK KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE KEYPAD WAS INTACT; NO DAMAGE WAS OBSERVED. THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS. THE COMPLAINT THAT THERE WERE LINES ON THE DISPLAY SCREEN WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(4).

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (LINE THROUGH DISPLAY) ISSUE. THE DISTRIBUTOR ALLEGED THAT THERE WERE LINES ON THE DISPLAY SCREEN. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.