FDA Adverse Event Malfunction Summary report: N

VENTED HUMIDIFICATION CHAMBER

MDR report key: 4120294 · Received September 26, 2014

Report

Report Number
9611451-2014-00782
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
August 27, 2014
Report Date
August 28, 2014
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290 CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR EVALUATION AND ARRIVED IN A CONDITION WHICH MADE IT IMPOSSIBLE FOR THE INVESTIGATION TO BE COMPLETED. RESULTS: THE CHAMBER HAD BEEN CRUSHED AND THE CHAMBER DOME WAS CRACKED. THE BASE OF THE DOME WAS DENTED AND WARPED. THE RT265 INFANT BREATHING CIRCUIT WAS ALSO DAMAGED. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 140507. CONCLUSION: THE DAMAGE TO THE COMPLAINT DEVICE WAS CAUSED BY TRANSPORTATION. DUE TO THE EXTENSIVE DAMAGE TO THE COMPLAINT DEVICE, WE WERE UNABLE TO PERFORM ANY TESTING AND THEREFORE WERE UNABLE TO DETERMINE WHAT CAUSED THE PROBLEM REPORTED BY THE CUSTOMER. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE THE FOLLOWING: - "SET APPROPRIATE VENTILATOR ALARM." - "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." FISHER & PAYKEL HEALTHCARE HAS CONTACTED THE TRANSPORTATION COMPANY TO ENQUIRE ON THE SOURCE OF THE DAMAGE WHICH PREVENTED THE DEVICE INVESTIGATION

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT AN MR290 HUMIDIFICATION CHAMBER WAS PRODUCING A RHYTHMIC WHISTLE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
602141 VENTED HUMIDIFICATION CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 140507

Patients

Seq Age Sex Outcome Treatment
1 DRÄGER INFINITY C500 VENTILATOR| FISHER & PAYKEL RT265 BREATHING CIRCUIT KIT