FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4120268 · Received September 26, 2014

Report

Report Number
2531779-2014-27564
Event Type
Malfunction
Date Received
September 26, 2014
Report Date
September 16, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 12/01/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/13/2014 WITH THE FOLLOWING FINDINGS: THE U-GROOVE WAS CRACKED ON THE PUMP AND THE BATTERY COMPARTMENT WAS ALSO CRACKED. THE THREADS ON THE BATTERY CAP WERE DAMAGED. THERE WAS EVIDENCE OF MOISTURE EVIDENT INSIDE OF BATTERY COMPARTMENT. THE PUMP FAILED LEAK TEST DUE TO THE BATTERY COMPARTMENT CRACKS. THERE WAS NO OTHER EVIDENCE OF MOISTURE PRESENT INSIDE OF BATTERY COMPARTMENT. UNRELATED TO THE BATTERY COMPARTMENT, WHEN THE PUMP WAS BOOTED THE DISPLAY IS A DIM RED COLOR.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.