ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-27486
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Report Date
- September 15, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING: VISUAL INSPECTION REVEALED THAT THE KEYPAD COVER WAS PEELING FORM THE BASE AND TORN BETWEEN THE UP ARROW AND CONTRAST BUTTON. DURING TESTING, THE UP ARROW KEYPAD BUTTON WAS FOUND TO BE UNRESPONSIVE. THE DOWN ARROW AND OK KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE OK KEYPAD BUTTON RESPONDED APPROPRIATELY TO BUTTON PRESSES. THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS. EVALUATION ALSO REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED. ADDITIONALLY, EVALUATION REVEALED THAT THE DISPLAY WAS DIM AND DISCOLORED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE UP ARROW KEYPAD BUTTON WAS UNRESPONSIVE WHILE THE DOWN ARROW AND OK KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. INVESTIGATION ALSO REVEALED THAT THERE WAS CONTAMINATION UNDER ALL KEYPAD BUTTON CONTACTS. FURTHERMORE, INVESTIGATION REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED. ADDITIONALLY, INVESTIGATION REVEALED THAT THE DISPLAY WAS DIM AND DISCOLORED. THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 602152 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR |