FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4120040 · Received September 25, 2014

Report

Report Number
2032227-2014-29364
Event Type
Malfunction
Date Received
September 25, 2014
Date of Event
August 28, 2014
Report Date
August 28, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP PASSED THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND NO DELIVERY TESTS. NO EXCESSIVE "NO DELIVERY" ALARM OR PRIME/FILL ANOMALY NOTED. THE DEVICE HAD CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND CRACKED RESERVOIR TUBE WINDOW.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED NO DELIVERY DURING A PRIME ATTEMPT. THE BLOOD GLUCOSE READING WAS 191 MG/DL. THE CUSTOMER STATED THAT SHE WAS ATTEMPTING TO DO A 5 UNIT PRIME BUT THE INSULIN PUMP DID NOT EXIT THE REWIND COMPLETE/BOLUS DELIVERY LOOP SUCCESSFULLY. SHE WAS UNABLE TO FILL THE TUBING. ADVISED REPLACEMENT OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
597892 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523NAS

Patients

Seq Age Sex Outcome Treatment
1 62 YR