FDA Adverse Event Malfunction Summary report: N

INTRA LUX HEAD 68 LH

MDR report key: 4118997 · Received September 19, 2014

Report

Report Number
3003637274-2014-00031
Event Type
Malfunction
Date Received
September 19, 2014
Date of Event
August 1, 2014
Report Date
August 26, 2014
Manufacturer
KAVO DENTAL GMBH
Product Code
EFB
PMA / PMN Number
K073478
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE ANALYSIS SHOWED THAT THE PUSH BUTTON OF THE CHUCK RELEASE STUCK IN. THIS CAUSED UNUSUAL INNER FRICTION AND THEREFORE HEAT UP OF THE HEAD. ROOT CAUSE WAS A DENT IN THE HEAD WHICH INDICATES THAT IT RECEIVED A STRONG HIT, E.G. DROPPING DURING RE-PROCESSING.

Description of Event or Problem · 1

DURING A STANDARD DENTAL TREATMENT THE HEAD HEATED UP AND CAUSED A BURN IN PATIENTS MOUTH. THE PATIENT WAS GIVEN AN OTC OINTMENT FOR THE BURN. IT HEALED ALREADY FULLY. FURTHER INFORMATION RELATED TO PATIENT HAVE NOT BEEN SUPPLIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
583763 INTRA LUX HEAD 68 LH DENTAL HANDPIECE EFB KAVO DENTAL GMBH 68 LH

Patients

Seq Age Sex Outcome Treatment
1 40 YR