ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-27391
- Event Type
- Malfunction
- Date Received
- September 24, 2014
- Report Date
- September 11, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE DEVICE WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/24/2014 WITH THE FOLLOWING FINDINGS: THE COMPLAINT COULD NOT BE DUPLICATED OR CONFIRMED WITH INVESTIGATION. REVIEW OF THE BLACK BOX REVEALED THE PUMP WAS MANUALLY SUSPENDED AT 07:34 AND RESUMED AT 08:24 ON (B)(6) 2014. BASAL HISTORY AND TOTAL DAILY DOSE BASAL HISTORY MATCHED THE USER PROGRAMMED BASAL RATES. THERE WAS NO ABNORMAL BEHAVIOR INDICATED IN THE BLACK BOX. THE PUMP SUCCESSFULLY COMPLETED A PRIME SEQUENCE, BOLUS DELIVERIES, AND 24-HOUR EXERCISE TEST WITHOUT ISSUE OR ALARM. THE PUMP PASSED A DELIVERY ACCURACY TEST. ALL DELIVERIES MADE DURING INVESTIGATION WERE ACCURATELY RECORDED IN THE BLACK BOX AND PUMP HISTORY.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THE PUMP¿S BASAL HISTORY DID NOT MATCH ACTIVE BASAL PROGRAM. IT WAS REPORTED THE PATIENT¿S BLOOD GLUCOSE (BG) WAS BETWEEN 250-499 MG/DL WITH NO SIGNS OR SYMPTOMS OF HYPERGLYCEMIA. THE REPORTED BG EXCURSION DOES NOT QUALIFY AS A SERIOUS INJURY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 594588 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR |