FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4113855 · Received September 24, 2014

Report

Report Number
2531779-2014-27299
Event Type
Malfunction
Date Received
September 24, 2014
Report Date
September 12, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/12/2014 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE KEYPAD IS CUT AND TORN UNDER THE DOWN KEY. A DAMAGED KEYPAD WILL ALLOW CONTAMINATION TO PERMEATE THE BUTTON CONTACTS NEGATIVELY IMPACTING THE BUTTON FUNCTION. THE DAMAGED KEYPAD IS OBVIOUS AND DETECTABLE AND SHOULD ALERT THE USER TO DISCONTINUE USE OF THE PUMP. EVALUATION REVEALED THAT ALL OF THE KEYS RESPOND APPROPRIATELY AND THERE WAS NO CONTAMINATION UNDER THE BUTTONS. THE DISPLAY WAS FOUND TO BE DIM AND PINK.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. EVALUATION REVEALED THAT THE DISPLAY WAS DIM AND PINK. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 09/12/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
594375 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 20 YR