SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Report
- Report Number
- 1416980-2014-32977
- Event Type
- Malfunction
- Date Received
- September 24, 2014
- Date of Event
- August 30, 2014
- Report Date
- August 30, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS IS A REPORT OF A SYSTEM ERROR 2240 ALARM THAT WAS CAUSED BY AN OPEN CLAMP ON AN UNUSED SUPPLY LINE. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE WARNS THE USER TO MAKE SURE ALL CLAMPS ON UNUSED FLUID LINES ARE CLOSED SECURELY AND INSTRUCTS THE USER TO CLOSE ALL CLAMPS WHILE PREPARING TO LOAD THE DISPOSABLE SET. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD RELEVANT ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SYSTEM ERROR 2240 (AIR IN SET) ALARM OCCURRED ON A HOME CHOICE DEVICE DURING DWELL ONE OF PERITONEAL DIALYSIS THERAPY. THE PATIENT WAS CONNECTED TO THE HOMECHOICE AT THE TIME OF THE ALARM. DURING THE TROUBLESHOOTING, IT WAS DETERMINED THAT THE PATIENT LEFT A CLAMP OPEN ON AN UNUSED SUPPLY LINE DURING THERAPY. THE POWER WAS CYCLED TO CLEAR THE ALARM AND PROPER PROCEDURES WERE REVIEWED WITH THE PATIENT. THE PATIENT WAS ADVISED TO START THERAPY OVER USING NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 595110 | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | HOMECHOICE |