FDA Adverse Event Death Summary report: N

MAQUET CARDIOHELP

MDR report key: 4113768 · Received September 3, 2014

Report

Report Number
8010762-2014-00324
Event Type
Death
Date Received
September 3, 2014
Date of Event
July 31, 2014
Report Date
August 5, 2014
Manufacturer
MAQUET CARDIOPULMONARY AG
Product Code
DTQ
PMA / PMN Number
K102726
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). MAQUET PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS AND RESOLUTION FOR THE DEVICE DESCRIBED IN THIS REPORT. A MAQUET SERVICE TECHNICIAN EXPLAINED VIA PHONE CALL WITH A NURSE, AN ENGINEER, AND THE FLIGHT NURSE, THAT IF THEY TURNED OFF THE GAS FLOW, THEN THE OXYGENATOR WOULD NOT RECEIVE ANY GAS AND IT WOULDN'T MATTER WHAT THE FIO2 SETTING WAS, THERE WAS NO GAS FLOW TO ALLOW ANY VENTILATION TO THE OXYGENATOR. AFTER THIS CONVERSATION THEY SEEMED TO UNDERSTAND THAT THE EVENT WAS RELATED TO HOW A BLENDER WORKS AND NOT THE CARDIOHELP FUNCTION. THE CARDIOHELP DEVICE HAS BEEN REQUESTED FOR RETURN TO PERFORM EVAL. A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED IF ADDITIONAL INFO BECOMES AVAILABLE. (B)(4).

Description of Event or Problem · 1

ORIGINALLY IT WAS REPORTED THAT NO ERROR ON A CARDIOHELP WAS NOTICED WHILE TRANSPORTING A PT. THEN IT WAS REPORTED THAT WHILE TRANSPORTING A PT ON A PLANE, THERE WAS A NO FLOW SITUATION REGARDING A SECHRIST AIR OXYGEN MIXER (BLENDER) BEING USED IN CONJUNCTION WITH A CARDIOHELP DEVICE. ONE OF THE TWO AIR SUPPLY PUMPS QUIT WORKING AND THE BLENDER BLEED-OFF ALARM ACTIVATED. SOMEONE THEN SELECTED 100% FIO2 WHILE THEY SHUT OFF THE BLENDER FLOW METER. IT WAS THEN NOTED THAT THE BLOOD GOT DARK AND THE PT CODED. ADDITIONAL INFO PROVIDED ON (B)(6) 2014 - DURING PT TRANSPORT THE BUBBLE SENSOR CLAMP POPPED OPEN AND THE RPM INDICATOR PORTION OF THE SCREEN BLINKED ON AND OFF. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
536664 MAQUET CARDIOHELP CARDIOHELP DTQ MAQUET CARDIOPULMONARY AG 70104.8012

Patients

Seq Age Sex Outcome Treatment
1 Death