MAQUET CARDIOHELP
Report
- Report Number
- 8010762-2014-00324
- Event Type
- Death
- Date Received
- September 3, 2014
- Date of Event
- July 31, 2014
- Report Date
- August 5, 2014
- Manufacturer
- MAQUET CARDIOPULMONARY AG
- Product Code
- DTQ
- PMA / PMN Number
- K102726
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). MAQUET PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS AND RESOLUTION FOR THE DEVICE DESCRIBED IN THIS REPORT. A MAQUET SERVICE TECHNICIAN EXPLAINED VIA PHONE CALL WITH A NURSE, AN ENGINEER, AND THE FLIGHT NURSE, THAT IF THEY TURNED OFF THE GAS FLOW, THEN THE OXYGENATOR WOULD NOT RECEIVE ANY GAS AND IT WOULDN'T MATTER WHAT THE FIO2 SETTING WAS, THERE WAS NO GAS FLOW TO ALLOW ANY VENTILATION TO THE OXYGENATOR. AFTER THIS CONVERSATION THEY SEEMED TO UNDERSTAND THAT THE EVENT WAS RELATED TO HOW A BLENDER WORKS AND NOT THE CARDIOHELP FUNCTION. THE CARDIOHELP DEVICE HAS BEEN REQUESTED FOR RETURN TO PERFORM EVAL. A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED IF ADDITIONAL INFO BECOMES AVAILABLE. (B)(4).
ORIGINALLY IT WAS REPORTED THAT NO ERROR ON A CARDIOHELP WAS NOTICED WHILE TRANSPORTING A PT. THEN IT WAS REPORTED THAT WHILE TRANSPORTING A PT ON A PLANE, THERE WAS A NO FLOW SITUATION REGARDING A SECHRIST AIR OXYGEN MIXER (BLENDER) BEING USED IN CONJUNCTION WITH A CARDIOHELP DEVICE. ONE OF THE TWO AIR SUPPLY PUMPS QUIT WORKING AND THE BLENDER BLEED-OFF ALARM ACTIVATED. SOMEONE THEN SELECTED 100% FIO2 WHILE THEY SHUT OFF THE BLENDER FLOW METER. IT WAS THEN NOTED THAT THE BLOOD GOT DARK AND THE PT CODED. ADDITIONAL INFO PROVIDED ON (B)(6) 2014 - DURING PT TRANSPORT THE BUBBLE SENSOR CLAMP POPPED OPEN AND THE RPM INDICATOR PORTION OF THE SCREEN BLINKED ON AND OFF. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 536664 | MAQUET CARDIOHELP | CARDIOHELP | DTQ | MAQUET CARDIOPULMONARY AG | 70104.8012 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Death |