FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 4113638 · Received September 17, 2014

Report

Report Number
2518422-2014-01581
Event Type
Malfunction
Date Received
September 17, 2014
Date of Event
August 26, 2014
Report Date
August 26, 2014
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A FAILURE OF THE DEVICE TO POWER ON WAS OBSERVED. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
575796 TRILOGY 100 CBK RESPIRONICS INC. 1054655

Patients

Seq Age Sex Outcome Treatment
1