CAD II
Report
- Report Number
- 8030965-2014-01388
- Event Type
- Malfunction
- Date Received
- September 24, 2014
- Report Date
- April 1, 2013
- Manufacturer
- SYNTHES GMBH
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL NARRATIVE: SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. (B)(6). DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE MOTOR SEIZED AND JAMMED, THERE WAS LOW TORQUE AND AN UNUSUAL NOISE. THIS WAS ATTRIBUTED TO NORMAL WEAR. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER ON (B)(4) 2013. PLACEHOLDER.
IT WAS REPORTED THAT THE TORQUE WAS TOO LOW AND THERE WAS AN ABNORMAL NOISE. THIS IS REPORT 1 OF 1 FOR COMPLAINT #(B)(4).