FDA Adverse Event Malfunction Summary report: N

PROXIMATE CIRCULAR STAPLER

MDR report key: 4113030 · Received September 24, 2014

Report

Report Number
3005075853-2014-06633
Event Type
Malfunction
Date Received
September 24, 2014
Date of Event
September 3, 2014
Report Date
September 4, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K940967
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS RESULTS FOUND THAT THE ECS29A DEVICE ARRIVED WITH THE ANVIL MISSING, OTHERWISE IN GOOD VISUAL CONDITION. AS THE ORIGINAL ANVIL WAS NOT RECEIVED, FURTHER INVESTIGATION WITH THE ORIGINAL ANVIL COULD NOT BE PERFORMED. THE BREAKAWAY WASHER WAS NOT PRESENT AND THERE WERE NO STAPLES PRESENT. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY WITH A TEST ANVIL; IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. IT SHOULD BE NOTED THAT WHEN REMOVING THE DEVICE; OPEN THE INSTRUMENT BY TURNING THE ADJUSTING KNOB COUNTERCLOCKWISE. FOR EASY REMOVAL, ONLY OPEN THE INSTRUMENT ONE-HALF TO THREE-FOURTHS REVOLUTIONS. TO ASSURE THE ANVIL IS FREE FROM TISSUE, ROTATE THE INSTRUMENT 90° IN BOTH DIRECTIONS. TO WITHDRAW THE OPEN INSTRUMENT, GENTLY APPLY REARWARD TRACTION WHILE SIMULTANEOUSLY ROTATING. PLEASE REFERENCE INSTRUCTIONS FOR USE FOR ADDITIONAL INFORMATION NEEDED. A BATCH RECORD REVIEW WAS PERFORMED AND THE BATCH HAD NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A COLON PROCEDURE, WHEN TRYING TO REMOVE AFTER THE ANASTOMOSIS, THE DEVICE WOULD NOT DISENGAGE. THE USER HAD TO RESECT THE ANASTOMOSIS FOR REMOVAL. THE AREA WAS OVER SEWED TO COMPLETE THE PROCEDURE. A LEAK WAS DETECTED AND THE AREA WAS OVER SEWED AGAIN TO MANAGE THE LEAK. THERE WAS NO PATIENT CONSEQUENCE OR CHANGE IN THE POST-OPERATIVE CARE OF THE PATIENT. NO ADDITIONAL INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
593241 PROXIMATE CIRCULAR STAPLER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA L4EY0J

Patients

Seq Age Sex Outcome Treatment
1