FDA Adverse Event
Injury
Summary report: N
PENTA
MDR report key: 4112936
·
Received September 17, 2014
Report
- Report Number
- 1627487-2014-08289
- Event Type
- Injury
- Date Received
- September 17, 2014
- Date of Event
- August 28, 2014
- Report Date
- August 28, 2014
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- P010032
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
Description of Event or Problem · 1
IT WAS REPORTED, THE PT WAS UNABLE TO INCREASE HIS SCS SYSTEM STIMULATION. AN SJM REPRESENTATIVE MET WITH THE PT FOR SYSTEM INTERROGATION AND DETERMINED THE PT'S SCS LEADS HAD INVALID IMPEDANCES. X-RAYS TAKEN SHOW THE PT'S LEAD HAD FRACTURED. SURGICAL INTERVENTION MAY TAKE PLACE AT A LATER DATE TO ADDRESS THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 577064 | PENTA | SCS LEAD | GZB | ST. JUDE MEDICAL - NEUROMODULATION | 3228 | 3582909 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | Other | SCS IPG: MODEL 3788| IMPLANTED: |