FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 4112830 · Received September 24, 2014

Report

Report Number
1416980-2014-32972
Event Type
Malfunction
Date Received
September 24, 2014
Date of Event
August 31, 2014
Report Date
August 31, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE PERFORMED. THE DAMAGED LATCH ROLLER WAS IDENTIFIED DURING THE VISUAL INSPECTION. THE CAUSE OF THE CONDITION COULD NOT BE DETERMINED. TO CORRECT THE CONDITION, THE LATCH ROLLER WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION, A FLOGARD INFUSION PUMP WAS FOUND WITH A DAMAGED LATCH ROLLER. THERE WAS NO REPORT OF PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
596163 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1