ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-27268
- Event Type
- Malfunction
- Date Received
- September 23, 2014
- Report Date
- September 11, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON 11/05/2014 WITH THE FOLLOWING FINDINGS: EVIDENCE OF UNCONFIRMED ALARMS AND PARTIALLY DISCHARGED BATTERIES BEING INSTALLED FOR USE WAS OBSERVED IN THE BLACK BOX DATA; USER ERROR CAUSED THE REPORTED BATTERY-LIFE ISSUE. THE BATTERY COMPARTMENT WAS FOUND TO BE INTACT. THE PUMP POWERED ON WITH THE RETURNED BATTERY CAP, WHICH WAS ABLE TO SECURE TO THE PUMP PER THE INSTRUCTIONS FOR USE (IFU). EVALUATION REVEALED THAT THE PUMP DREW ELECTRIC CURRENT AT LEVELS WITHIN THE SPECIFICATIONS: THE REPORTED BATTERY-LIFE ISSUE WAS NOT DUPLICATED. THE PUMP CASE WAS REMOVED, AND NO EVIDENCE OF INTERNAL DAMAGE OR DEFECT WAS FOUND. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 590105 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |