ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-27248
- Event Type
- Malfunction
- Date Received
- September 23, 2014
- Report Date
- September 11, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/17/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP BLACK BOX DATA REVEALED MULTIPLE UNEXPLAINED PUMP REBOOTS. DURING A VISUAL INSPECTION OF THE PUMP, NO DAMAGE WAS FOUND TO THE BATTERY COMPARTMENT OR RETURNED BATTERY CAP. THE RETURNED BATTERY CAP SECURED PROPERLY TO THE PUMP, AND A POWER LOSS WAS NOT OBSERVED. THE BATTERY CAP CONTACT DIMENSIONS MEASURED WITHIN SPECIFICATIONS. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER INTERRUPTIONS. THE PUMP CASE WAS REMOVED, AND NO EVIDENCE OF MOISTURE INGRESS OR DAMAGE TO THE POWER CIRCUIT WAS FOUND. UNRELATED TO THE COMPLAINT, THE PUMP WAS POWERED ON AND THE DISPLAY SCREEN APPEARED DIM AND DISCOLORED.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 591527 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR |