FDA Adverse Event Malfunction Summary report: N

ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 4112329 · Received September 23, 2014

Report

Report Number
9611451-2014-00779
Event Type
Malfunction
Date Received
September 23, 2014
Date of Event
August 26, 2014
Report Date
September 1, 2014
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
K983112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT RT340 BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (FPH) IN (B)(4) AND WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT THE EVAQUA EXPIRATORY LIMB HAD A HOLE ABOUT 26CM FROM THE PROXIMAL CONNECTOR. CONCLUSION: BASED ON OUR VISUAL INSPECTION, THE SUBJECT EXPIRATORY LIMB APPEARED TO HAVE BEEN PUNCTURED BY A BLUNT OBJECT. ALL RT340 BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED FOR LEAKS BEFORE RELEASING FOR DISTRIBUTION. ANY BREATHING CIRCUIT WHICH FAILS IS DISCARDED. IN ADDITION, TUBE WEIGHING AND BOND STRENGTH TESTING ARE PERFORMED EVERY 15 MINUTES. IF ANY FAULTS ARE DETECTED THE WHOLE BATCH IS PLACED ON HOLD FOR INVESTIGATION. THIS SUGGESTS THAT THE SUBJECT BREATHING CIRCUIT WAS DAMAGED AFTER IT WAS RELEASED FOR DISTRIBUTION. THE KEY DIFFERENCE BETWEEN THE EVAQUA BREATHING CIRCUITS AND CONVENTIONAL BREATHING CIRCUITS IS THAT THE EXPIRATORY LIMB OF THE EVAQUA CIRCUITS SUCH AS THE RT340 IS COMPOSED OF A THIN, SEMI-PERMEABLE FILM SPECIALLY DESIGNED TO ALLOW WATER VAPOUR FROM EXPIRED VENTILATORY GAS TO PASS THROUGH. THE EVAQUA EXPIRATORY TUBE HAS A PROTECTIVE MESH WHICH PREVENTS DAMAGE TO THE WALLS OF THE TUBE, HOWEVER THE EVAQUA TUBING CAN BE MORE SUSCEPTIBLE TO DAMAGE THAN CONVENTIONAL CIRCUITS WHEN EXPOSED TO ROUGH HANDLING OR DAMAGE CAUSED BY SHARP OBJECTS AND NON-FPH CIRCUIT HANGERS. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT. SET APPROPRIATE VENTILATOR ALARMS. FIT ONLY THE SUPPLIED FISHER & PAYKEL HEALTHCARE CIRCUIT HANGER WITH CARE TO AVOID CIRCUIT DAMAGE. HOSPITAL STAFF CORRECTLY CHECKED THE BREATHING CIRCUIT BEFORE PATIENT USE, WHICH IS IN LINE WITH OUR USER INSTRUCTIONS.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT AN RT340 ADULT BREATHING CIRCUIT FAILED THE PB840 VENTILATOR LEAK TEST AND THAT THEY FOUND A PIN HOLE ON THE EXPIRATORY LIMB. THIS WAS FOUND BEFORE PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
591463 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT340

Patients

Seq Age Sex Outcome Treatment
1