FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAL

MDR report key: 4112161 · Received September 23, 2014

Report

Report Number
2032227-2014-27561
Event Type
Malfunction
Date Received
September 23, 2014
Date of Event
August 25, 2014
Report Date
August 25, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO BUTTON ERROR ALARM WAS NOTED. THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED RESERVOIR TUBE, A STAINED END CAP STICKER AND CRACKED BATTERY TUBE THREADS. NO MOISTURE DAMAGE WAS NOTED ON THE ELECTRONICS, MOTOR, VIBRATOR MOTOR OR BATTERY TUBE ASSEMBLY.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR AND WAS NOT RESPONDING. CUSTOMER STATED THAT SHE WAS AT PRACTICE AND WAS WEARING THE INSULIN PUMP IN HER SPORTS BRA. INSULIN PUMP MAY HAVE BEEN EXPOSED TO SWEAT. CUSTOMER WAS UNABLE TO CLEAR THE ALARM. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
592084 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAL INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523NAL

Patients

Seq Age Sex Outcome Treatment
1 12 YR