PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-28087
- Event Type
- Malfunction
- Date Received
- September 23, 2014
- Date of Event
- August 25, 2014
- Report Date
- August 25, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK
- Reporter Occupation
- PATIENT
Narratives
THE INSULIN PUMP PASSED THE DISPLACEMENT, BASIC OCCLUSION, AND PRIME TESTS. HOWEVER, THE INSULIN PUMP WAS RECEIVED STUCK IN A MOTOR ERROR ALARM DURING BOLUS DELIVERY LOOP. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING OUR TESTING. NO ERROR ALARM WAS CONFIRMED DUE TO ERASED HISTORY FILE. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, CRACKED RESERVOIR TUBE LIP, AND A BROKEN BELT CLIP SLOT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE FILL TUBING PROCEDURE, AS WELL AS ANOTHER ERROR ALARM. THE CUSTOMER DID NOT KNOW THE BLOOD GLUCOSE READING. ADVISED REPLACEMENT OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.