PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4112097 · Received September 23, 2014

Report

Report Number
2032227-2014-27492
Event Type
Malfunction
Date Received
September 23, 2014
Date of Event
August 25, 2014
Report Date
August 25, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM, NUMBER RAMPING, OR SCROLL BAR ANOMALY WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, CRACKED RESERVOIR TUBE LIP, AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD AN UNRESPONSIVE KEYPAD AND BUTTONS; THE BUTTONS KEPT STICKING. AND THE NUMBERS WERE SCROLLING. THE CUSTOMER STATED THAT USUALLY THEY HAVE THE DEVICE CLIPPED TO THEIR SHIRT OR AROUND THEIR NECK. THE CUSTOMER'S BLOOD GLUCOSE WAS 147 MG/DL. THE CUSTOMER STATED THAT THEY WERE IN THE HOSPITAL FOR ISSUES WITH THEIR LVAT MACHINE, NON DIABETES RELATED EVENT. THE DISCONTINUATION OF THE INSULIN PUMP ADVISED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
591919 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 59 YR