FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4111767 · Received September 23, 2014

Report

Report Number
2032227-2014-28014
Event Type
Malfunction
Date Received
September 23, 2014
Date of Event
August 25, 2014
Report Date
August 25, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED STUCK IN MOTOR ERROR LOOP DURING BOLUS AND BASAL DELIVERY. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED; THE MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING OUR TESTING. HOWEVER, THE INSULIN PUMP PASSED DISPLACEMENT, REWIND, BASIC OCCLUSION AND PRIME/A33 TESTS. THE INSULIN PUMP HAS BATTERY TUBE THREADS CRACKED, MINOR SCRATCHED LCD WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER AND STAINED END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).

Description of Event or Problem · 1

IT IS REPORTED THAT CUSTOMER EXPERIENCED MOTOR ERROR ALARM. INSULIN PUMP WILL BE REPLACED. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
592296 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-554WWS

Patients

Seq Age Sex Outcome Treatment
1