FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4111452 · Received September 23, 2014

Report

Report Number
2032227-2014-27999
Event Type
Malfunction
Date Received
September 23, 2014
Date of Event
August 24, 2014
Report Date
August 25, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE BASIC OCCLUSION TEST AND THE DISPLACEMENT TEST. NO MOTOR ERROR ALARMS WERE NOTED. HOWEVER, THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST DUE TO A FAULTY FORCE SENSOR RESISTOR. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE INSULIN PUMP WAS ALSO RECEIVED WITH SOME COSMETIC DAMAGE.

Description of Event or Problem · 1

IT IS REPORTED THAT THE PATIENT'S SENSOR INSULIN PUMP DISPLAYED A MOTOR ERROR. THE PATIENT'S BLOOD GLUCOSE LEVEL WAS UNKNOWN. TROUBLE SHOOTING WAS CONDUCTED AND WE EXPLAINED THE POSSIBLE CAUSES FOR THE ALARM. DEVICE IS BEING RETURNED. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
590122 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-754WWL

Patients

Seq Age Sex Outcome Treatment
1 114 YR