BALL HEX SCREWDRIVER 8MM FOR TI FEMORAL NAILS
Report
- Report Number
- 2530088-2014-10285
- Event Type
- Malfunction
- Date Received
- September 23, 2014
- Date of Event
- August 26, 2014
- Report Date
- August 27, 2014
- Manufacturer
- SYNTHES BRANDYWINE
- Product Code
- HXX
- PMA / PMN Number
- PEXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A MANUFACTURING INVESTIGATION WAS CONDUCTED. THE REPORT INDICATES THAT NO COMPLAINT IS FOR OILY RESIDUE FOUND. NO RESIDUE CAN BE FOUND ON THE RETURNED DEVICE. BECAUSE NO RESIDUE IS PRESENT THIS COMPLAINT IS UNCONFIRMED. DECONTAMINATION WAS PERFORMED PRIOR TO THIS MANUFACTURING EVALUATION AND THE DECONTAMINATION PROCESS MAY HAVE COMPROMISED THE FOREIGN SUBSTANCE. BECAUSE THE RESIDUE IS NOT PRESENT FURTHER TESTING WILL NOT BE PERFORMED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
A PRODUCT DEVELOPMENT INVESTIGATION WAS CONDUCTED. THE TIP OF THE SCREWDRIVER IS WORN, AND THE HANDLE HAS VARIOUS MARKS AND SCRATCHES. NO RESIDUE WAS NOTED ON THE RETURNED DEVICE. THE RETURNED DEVICES SHOW SIGNS OF ROUTINE USE DURING THEIR LIFESPANS. THE COMPLAINT CONDITION HAS TO DO WITH A RESIDUE ON THE HANDLES; BUT DUE TO DECONTAMINATION UPON RECEIPT OF THE COMPLAINT PARTS, THERE WAS NO RESIDUE PRESENT ON THE DEVICES. THUS THE COMPLAINT CONDITION IS UNCONFIRMED, AND WAS MOST LIKELY DUE TO REPEATED STERILIZATION CYCLES, WITH A LUBRICANT ADDITIVE PRESENT, OVER THE LIFE OF THE INSTRUMENTS. NO DRAWING ISSUES OR DISCREPANCIES WERE NOTED. THE DESIGN IS ADEQUATE FOR ITS INTENDED USE WHEN USED AND MAINTAINED AS RECOMMENDED AND DID NOT CONTRIBUTE TO THIS COMPLAINT CONDITION. NO DESIGN DEFICIENCIES WERE DISCOVERED DURING THE INVESTIGATION. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
ADDITIONAL NARRATIVE: THE DEVICE WAS RECEIVED, THE INVESTIGATION COULD NOT BE COMPLETED, AND NO CONCLUSION COULD BE DRAWN, AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL NARRATIVE: THE DEVICE HISTORY RECORD WAS REVIEWED AND MATERIAL REVIEW REPORT ( MRR) #(B)(4) WAS FOUND ON P/N 357.515.01 LOT #4895281 FOR SPOTS/UNEVEN FINISH ON 82 OUT OF 120 PARTS. THE NONCONFORMING PARTS WERE REWORKED, INSPECTED AND PASSED. THIS NONCONFORMANCE IS NOT RELEVANT TO THE COMPLAINT CONDITION BECAUSE THE ISSUE IS VISUAL ON THE BALL HEX SCREWDRIVER COMPONENT NOT FINISHED PRODUCT. NO ISSUES WERE FOUND DURING MANUFACTURE THAT WOULD CONTRIBUTE TO THE COMPLAINT CONDITION. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. (B)(6). THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. A REVIEW OF THE DEVICE HISTORY RECORDS HAS BEEN REQUESTED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
SURGERY WAS DELAYED AND RE-SCHEDULED FOR (B)(6) 2014. RE-SCHEDULING OF SURGERY DID NOT CAUSE ANY ADVERSE EVENT TO PATIENT.
DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THE INSTRUMENTS IN A TROCHANTERIC FIXATION NAIL (TFN) SET WERE FOUND TO HAVE A CLEAR OILY RESIDUE DEPOSITED ON THE TRAYS UNDER THE SCREWDRIVER HANDLES. THIS WAS DISCOVERED PRIOR TO USE WHEN OPENING THE SETS FOR USE IN A PROCEDURE, NO PATIENT INVOLVEMENT AS THE SURGERY WAS RE-SCHEDULED. THIS IS REPORT 6 OF 7 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 589673 | BALL HEX SCREWDRIVER 8MM FOR TI FEMORAL NAILS | GUIDE | HXX | SYNTHES BRANDYWINE | 4936924 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |